Single channel Preamplifier/Discriminator PAD01S
Home Introduction CEM Models CEM Handling Quality Assurance Preamplifier Measurement Systems

 

 

 

Description

This low cost preamplifier is designed for pulse counting of typical input signals for channel electron multipliers (CEMs) or photo multiplier tubes (PMTs).

Theoretical maximum count rate is 10e7 pulses per second, dead time and output pulse width are entirely independent of threshold setting and overdrive, but change slightly (+/- 20% max) with rate. The primary application therefore is pulse counting at not too high rates (some 10e5/s).

 

  Case
Units
Range
  Aluminium (RF-schielded)
mm
90 x 35 x 31
  Output to be connected to
  Electronic counter
  Input signal (negative pulse)
  Amplitude into 50 Ohm
mV
- 3 to - 1000
  Safe operation without damage
V
+/- 3
  Pulse width (FWHM)
ns
5 to 20
  Count rate (Poission distribution)
cps
up to 5 x 10e5
  Resistance, dc-coupled
Ohm
50
  Threshold, screwdriver adjustable
mV
-3 to - 130
  Output signal (positive pulse)
  Resistance
Ohm
50
 
Amplitude into high Z load
V
2.2
 
Amplitude into 50 Ohm
V
1.1
  Pulse width
ns
40 +/- 10
  Dead time
ns
80 +/- 20
  Power Supply
  A power supply for 100 - 240 V, 50/60 Hz is delivered with the PAD01S
V
+6 (+/- 5 %), 24 mA typ.
 
V
- 6 (+/- 5 %), 14 mA typ.
  Temperature range
  For specified data
°C
10 - 40
  For operation
°C
0 - 70
       

 

 
Fig. 1 The preamplifier/discriminator WMT PAD01S
 
 
Fig. 2 The screwdriver adjustable threshold
 
 
 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 


Dr. Sjuts Optotechnik GmbH

Am Schlehdorn 1, 37077 Goettingen, Germany Tel. + 49 (0) 551 209 95 62, Fax. + 49 (0) 551 209 95 63